New TopShim run:

version = topshim 2.2.0
date = 2024-04-12
time = 00:49:56.912
command = topshim 

FINE SHIMMING

Parameters:

probe = Z163739_0737
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.7 Hz
Iteration 2: FWHM Before= 0.7 Hz, Target= 0.7 Hz, After= 0.6 Hz
Iteration 3: FWHM Before= 0.6 Hz, Target= 0.3 Hz, After= 0.6 Hz
Iteration 4: FWHM Before= 0.6 Hz, Target= 0.4 Hz, After= 0.6 Hz
Iteration 5: FWHM Before= 0.6 Hz, Target= 0.4 Hz, After= 0.6 Hz
Iteration 6: FWHM Before= 0.6 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration 7: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 0.3 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D shimming, non-spinning) = 0.636

shim changes: 
Z	-546
Z2	-139
Z3	-164
Z4	-341
Z5	-134
Z6	-1966
Z7	-4117
Z8	-874

duration = 4 min 13 sec
completed successfully
finished Fri Apr 12 00:54:07 2024


========================================================================